Click here to enlarge imageWastewater influent or industrial waste is pumped to a sample overflow chamber in excess. A subsample of the waste is pumped into a aerator and mixed with a readily biodegradable substrate and nutrient stock.
The substrate assures that the biomass will have sufficient food for vigorous oxygen demand. Air is introduced in the aerator to mix the sample and nutrients, as well as to increase the dissolved oxygen level to saturation. The ratio of wastewater to biodegradable substrate can be set by adjusting the rate of the individual feed pumps. The usual ratio is approximately 1 to 10. The mixture flows through the biological filter at a rate of 40 ml per minute. This rate allows the healthy biomass to consume most of the available food and dissolved oxygen.
If the wastewater contains materials that inhibit or kill the microbial inhibition screening culture in the filter, the microorganisms consume little or no oxygen. Thus, the quantity of dissolved oxygen (DO) passing out of the filter is a direct indication of inhibition or toxicity. The dissolved oxygen remaining in the effluent of the filter is monitored by the system's data recorder through a dissolved oxygen sensor. The results are continuously recorded by a computer. Alarms may be set through the computer to provide a local or a remote signal, such as on the central control panel.
Application Data
The high speed of the biological filter's reaction to inhibitory or toxic conditions, combined with the user selected alarm points, gives operating personnel adequate warning so that remedial action may be taken. In laboratory testing, response times ranged from 8 to 20 minutes with varying concentrations of sodium hypochlorite.
During testing with 100 ppm sodium hypochlorite, the Bioscan system reached the alarm point of 1.5 ppm dissolved oxygen within 10 minutes. A feed pump control is set at 3 ppm dissolved oxygen. When the dissolved oxygen exceeds this level, the feed pump is stopped. This protects the biological filter from excessive exposure to highly inhibitory waste streams by pausing the introduction of the offending sample. The nutrient input continues to dilute the material in the biological filter. When the filter's respiration recovers enough to drop the dissolved oxygen level below 2.9 ppm, the feed pump resumes.
Self-testing of the dissolved oxygen system is provided by the computer on a daily basis. The system drains the oxygen probe chamber and performs an air calibration every 24 hours. Sample pump delivery adjustments and dissolved oxygen system calibration can be done manually by plant operators without interfering with the computer settings. Completed daily records are stored in the data recorder's memory for up to four days. The computer automatically assigns the correct date to each chart. After four days the older data is over-written by current data.